SYSMEX CA-1500

SYSMEX CA-1500

Location:Przeźmierowo, Poland

Available quantity:1

Description

The housing was painted. The missing halogen bulb (JCR6V10W20H-SY) was supplemented. An alternative (made) waste box was added to the analyser. The waste box sensor was replaced with a working one (it was replaced with the additional feeder sensor). Fluid detection control (via needles, fill the A2 and E3 vessels 1.5 mm above the level of the trays) – correct. Hydraulics were drained (service program). The analyser starts as expected. The heating block and reagent cooling are working. It starts without error messages, the “Ready” label appears on the screen.


Specifications

ManufacturerSYSMEX
ModelCA-1500
ConditionUsed
Serial NumberA2566
Stock Number9163001248